Anaesthesiology

The Anesthesia Workstation and Delivery Systems for Inhaled Anesthetics

Built from Barash Clinical Anesthesia, 9th Edition

What’s inside

14 sections · 196 slides

  1. 01

    Overview

    • Scope of this topic
    • Importance of the gas delivery system

    2 slides

  2. 02

    Anatomy of the Delivery System

    • Basic configuration of the delivery system
    • The five stages of gas delivery
    • Function of the machine and the vaporizer
    • Function of the patient breathing circuit
    • Waste gas scavenging and the term workstation
    • Generic two-gas anesthesia machine
    • Three pressure circuits of the machine
    • Pressures in the high- and intermediate-pressure circuits
    • Contents of the low-pressure circuit
    • Two supply sources for every gas
    • Flow control valves and the path to the outlet
    • The outlet check valve and the oxygen flush connection

    12 slides

  3. 03

    Pipeline and Cylinder Supply

    • The hospital central piping system
    • Reported pipeline failures and crossover deaths
    • Management of a suspected pipeline crossover
    • Reason the pipeline must be disconnected
    • Gas-specific wall outlet connections
    • Wall outlets and DISS connections
    • Machine gas inlet and cylinder pin connections
    • Backup cylinders and the hanger yoke
    • The Pin Index Safety System and its limits
    • E-cylinder valve and yoke components
    • Cylinder check valves and pressure gauge
    • Cylinder pressure regulator and the transport valve
    • Rationale for keeping cylinder valves closed
    • Cylinder DISS outlet as an oxygen backup
    • Boyle's law and the volume in a full E-cylinder
    • Estimating remaining oxygen supply time
    • Physical properties of nitrous oxide
    • Contents of a nitrous oxide E-cylinder
    • Reading the nitrous oxide cylinder gauge
    • Nitrous oxide supply into the machine
    • Oxygen paths within the intermediate-pressure system

    21 slides

  4. 04

    Oxygen Supply Pressure Failure Devices

    • Cascade of protective devices and the pressure alarms
    • The oxygen failure cutoff valve
    • Pressure-sensor shutoff valve in the open and closed positions
    • Threshold valves and the second-stage oxygen regulator
    • Failure of the fail-safe valve to prevent a hypoxic mixture
    • Five ways a hypoxic mixture can still be delivered

    6 slides

  5. 05

    Flowmeters and Gas Flow Control

    • Oxygen flowmeter assembly
    • Thorpe tubes and electronic flow sensors
    • Electronic presentation of the flow rate
    • The annular space
    • Constant-pressure variable-orifice flowmeters
    • Flow tube constriction and the physics of gas flow
    • Flow control knob safety and the flowmeter scale
    • Electronic control panel with soft keys
    • Flow tubes, indicator floats and float stops
    • Flowmeter leaks and causes of inaccuracy
    • Flowmeter sequence as a cause of hypoxia
    • Oxygen flow tube leak
    • Electronic flow display and fresh gas flow optimization
    • Total flow indicator that survives a power failure
    • The GE Aisys Carestation gas mixer
    • Alternate oxygen flowmeter backup system
    • Low-flow wizard recommendations
    • How the Low Flow Wizard calculates, and its limits
    • The GE ecoFLOW tool

    19 slides

  6. 06

    Proportioning Systems

    • Purpose of nitrous oxide/oxygen proportioning systems
    • Ohmeda Link-25 Proportion-Limiting Control System
    • Mechanics and pneumatics of the Link-25 system
    • How the Link-25 corrects an unsafe setting
    • Draeger Sensitive Oxygen Ratio Controller
    • Components and operating principle of the S-ORC
    • Behaviour of the S-ORC shaft
    • Link-25 compared with the S-ORC
    • Wrong supply gas defeats both systems
    • Defective mechanics and leaks downstream
    • Oxygen dilution by inert gases and volatile agents
    • The oxygen flush valve

    12 slides

  7. 07

    Vaporizer Physics

    • Evolution and classification of vaporizers
    • Physical properties of potent inhaled volatile anesthetic agents
    • Saturated vapor pressure
    • Saturated vapor pressure versus temperature curves
    • Boiling point and latent heat of vaporization
    • Specific heat and thermal conductivity
    • Physical properties that shape vaporizer design

    7 slides

  8. 08

    Variable Bypass Vaporizers

    • Meaning of each classification term
    • Generic variable bypass vaporizer
    • Principle of dilution in a variable bypass vaporizer
    • Path of gas through the vaporizer
    • Filling the sump and estimating agent consumption
    • Worked example: 1% sevoflurane
    • Worked examples: 2% sevoflurane and isoflurane
    • Bimetallic strip temperature compensation
    • Expansion element temperature compensation
    • Ideal vaporizer performance and the effect of fresh gas flow
    • Temperature compensation and vaporizer safety features
    • The pumping effect
    • Proposed mechanism of the pumping effect
    • Limiting the pumping effect, and carrier gas solubility
    • Influence of carrier gas composition on delivered concentration
    • Misfilling and deliberate agent-vaporizer mismatch
    • Predicting the output of a misfilled vaporizer
    • Tipping of a vaporizer
    • Management after tipping, and improper filling
    • Tipping-resistant designs and use in the MRI suite
    • Vaporizer leaks and leak testing

    21 slides

  9. 09

    Vaporizers for Desflurane

    • Physical properties that force a new design
    • Unsuitability of variable bypass vaporizers for desflurane
    • Tec 6 desflurane vaporizer schematic
    • Heated sump, shutoff valve and pressure regulator
    • How the two circuits are matched
    • Working pressure of the Tec 6
    • Fresh gas flow rate versus working pressure
    • Fresh Gas Flow Rate versus Working Pressure
    • Constant fresh gas flow with a rising dial setting
    • Dial Setting versus Flow through Restrictor R2
    • Constant dial setting with a rising fresh gas flow
    • Effect of altitude on older Tec type vaporizers
    • Performance of older Tec type vaporizers versus the Tec 6 at varying ambient pressures
    • Partial pressure and ambient pressure compensation
    • Tec type versus Tec 6 output across ambient pressures
    • Altitude correction and carrier gas effects on the Tec 6
    • Desflurane misfilling hazard and the Saf-T-Fill adapter
    • Misfilling a Tec 6 and what triggers its shutoff valve
    • Desflurane vaporizer alarm indicators
    • Meaning of the D-Vapor alarms

    20 slides

  10. 10

    Aladin Cassette and Injection Vaporizers

    • Aladin cassette concept and injection vaporizers
    • Aladin sevoflurane cassette
    • Aladin cassettes mounted on a GE Aisys Carestation
    • Aladin cassette vaporizer schematic
    • Inputs to the Aladin central processing unit
    • Aladin check valve and agent heating
    • Aladin behaviour above the boiling point of desflurane
    • Comparison of vaporizer types
    • Vaporizer Models and Characteristics

    9 slides

  11. 11

    The Circle Breathing System

    • Function of the breathing circuit and recent design changes
    • Components of the circle breathing system
    • Seven primary components of the circle system
    • Universal F circuit and the three anti-rebreathing rules
    • Semiopen, semiclosed and closed circle systems
    • Efficiency and advantages of the circle system
    • Hazards of circle system complexity and obstruction
    • Malfunction of the unidirectional valves
    • Flow transducer leaks and kinked tubing
    • Coaxial versus dual-limb breathing tubes
    • Effect of longer breathing circuit tubing

    11 slides

  12. 12

    Carbon Dioxide Absorbents

    • Absorbent safety concerns and the ideal absorbent
    • Canister leaks and the cost of calcium hydroxide lime
    • Composition of high-moisture soda lime
    • Calcium hydroxide lime
    • Granule size and absorptive capacity
    • The chemistry of carbon dioxide absorption
    • Sequence of the absorption reaction
    • Ethyl violet as a pH indicator
    • Failure of the colour indicator
    • Compound A: formation and the factors that raise it
    • Effect of absorbent dehydration on compound A
    • Carbon monoxide production from desiccated absorbents
    • The Monday morning pattern of carbon monoxide poisoning
    • Retrograde drying and low-strong-base products
    • Factors that increase carbon monoxide production
    • Interventions to reduce carbon monoxide exposure
    • APSF 2005 recommendations on absorbent safety
    • Breathing system fires and lithium-based absorbents
    • Ingredients of a breathing system fire
    • Litholyme indicator and SpiraLith Ca
    • Absorbent comparisons

    21 slides

  13. 13

    Mapleson Circuits and the Resuscitation Bag

    • Origin and components of the Mapleson systems
    • Mapleson breathing systems
    • Functional groups within the Mapleson family
    • Flow requirement of the Mapleson A circuit
    • Rebreathing variables and flow needs of the Mapleson systems
    • Structure of the Bain circuit
    • Advantages and hazards of the Bain circuit
    • The Pethick test of inner tube integrity
    • The Jackson-Rees circuit
    • The self-inflating resuscitation bag as plan B

    10 slides

  14. 14

    Summary and Key Points

    • Key point: limits of the oxygen failure cutoff valves
    • Key point: two actions in a pipeline crossover
    • Key point: why desflurane needs a special vaporizer
    • Key point: absorbent interactions with inhaled anesthetics
    • Key point: the low-pressure circuit as the vulnerable area
    • Key point: which leak test to use
    • Summary: the journey of one breath
    • Summary: the recurring safety themes
    • Sources cited in this topic
    • Sources cited in this topic (continued)
    • Sources cited in this topic (continued)
    • Barash, Cullen, and Stoelting's Clinical Anesthesia, 9th Edition

    12 slides